1.
Journal of Korean Medical Science ; : e107-2018.
Article in English | WPRIM | ID: wpr-714282

ABSTRACT

BACKGROUND: To analyze clinical outcome of CyberKnife (CK) tumor-tracking stereotactic body radiotherapy (SBRT) for prostate cancer (Pca) according to the magnitude of intra-fractional prostate motion. METHODS: Medical records and daily treatment logs for 71 patients who received CK tumor-tracking SBRT were retrospectively analyzed. Statistical relationships between prostate motion and various outcome results, including local recurrence (LR), biochemical failure (BF), and treatment-related toxicity, were investigated in order to evaluate motion-dependent efficacy of tumor-tracking SBRT for Pca. RESULTS: In a total 71 patients, 3 (4.2%) patients with LR, 12 (16.9%) patients with BF, and 22 (31%) patients with grade-II or worse toxicities to rectal or bladder (22 to rectal, 22 to bladder and 8 patients to both) were observed in a median follow-up of 47 months. Magnitudes of intra-fractional tumor motion along superior-inferior, right-left, and anterior-posterior (AP) axes were 0.15 ± 0.31, 0.12 ± 0.19, and 0.73 ± 0.32 mm, respectively. Radial magnitude was estimated to be 1.0 ± 0.35 mm. Intra-fractional movement was not significantly correlated with tumor control. However, it was significant correlated with the incidence of grade-II or worse toxicity to rectum or bladder particularly when tumor motion was in the AP axis. CONCLUSION: Our quantitative results revealed that toxicity related to SBRT treatment was highly sensitive to intra-fractional prostate movements, although local-tumor control was not affected by such movements. Our results demonstrate that precise motion correction is essential in prostate SBRT, even if it seems to be small.


Subject(s)
Humans , Follow-Up Studies , Incidence , Medical Records , Passive Cutaneous Anaphylaxis , Prostate , Prostatic Neoplasms , Radiosurgery , Rectum , Recurrence , Retrospective Studies , Urinary Bladder
2.
Korean Journal of Medical Physics ; : 123-126, 2012.
Article in Korean | WPRIM | ID: wpr-95725

ABSTRACT

For the purpose of quality assurance (QA) of the blood irradiator, QA programs for daily, monthly, and yearly were developed. For daily tests, simple items for basically operating the machine are recommended. For monthly and yearly tests, the measurement of dose to assure the dose delivery system are performed by a dosimetry devices (Glass dosimeter jig) developed in this study. The QA program is practical for clinical environment.


Subject(s)
Glass
3.
Korean Journal of Medical Physics ; : 124-130, 2011.
Article in Korean | WPRIM | ID: wpr-99721

ABSTRACT

In this study the dosimetric evaluation for a biological sample irradiated by gamma rays from Cs-137 irradiator (Gamma Irradiator, Chiyoda Technol Co., Japan) was performed for radiobiological experiment. A spherical water with a diameter of 3 cm was assumed as a biological sample. The absorbed dose were determined by the air kerma based dosimetric calculation system. The theoretical and Monte Carlo calculations (MCNPX) were performed and compared to evaluate measured air kerma and determined absorbed dose respectively. As a result of comparison with theoretical calculation, the measured air kerma was in good agreement within 3.1% at the distance of 100 and 200 cm from the source. In comparison with Monte Carlo results the determined absorbed dose along the central axis was in good agreement within 1.9% and 3.7% at 100 cm and 200 cm respectively. Although the preliminary results were obtained in this study these results were used as a basis of dosimetric evaluation for radiobiological experiment. Extended study will be performed to evaluate the dose in various conditions of biological samples.


Subject(s)
Axis, Cervical Vertebra , Gamma Rays , Water
4.
Korean Journal of Medical Physics ; : 148-154, 2011.
Article in Korean | WPRIM | ID: wpr-99718

ABSTRACT

The quality factors (kQ,Q0) were evaluated by appling the results recently studied for the effect of central electrode in TRS-398 protocol. The PTW-31010 and IBA-CC13 chambers were used in this study. The quality factors were calculated as a function of beam quality for high energy electron and photon beams and compared with data currently used in TRS-398 protocol. In the PTW-31010 chamber using aluminium electrode, appling the new central electrode collections, the quality factors were 0.4% and 0.9% higher than current TRS-398 data for high energy photon and electron beams respectively. In the IBA-CC13 chamber using C-552 electrode, there are no variations in quality factors compared to TRS-398 data currently used.


Subject(s)
Electrodes , Electrons
5.
Korean Journal of Medical Physics ; : 78-85, 2010.
Article in Korean | WPRIM | ID: wpr-30101

ABSTRACT

The compatibility with GammaMed-12i brachytherapy machine for an Ir-192 encapsulated source (IRRS20, KAERI, Korea) manufactured by Korea atomic energy research institute (KAERI) has been investigated. As a mechanical testing of compatibility, precise measurement of step movement with channels, measurement of curvature of radius for wire, and emergency return testing were performed. Periodic measurements of air kerma strength for 45 days were carried out to evaluate decay characteristics of Ir-192 radioisotope and comparison of dose distributions in phantom between KAERI and old sources previously used were performed by film dosimetry. KAERI source has a good compatibility with GammaMed12i machine as a result of mechanical testing. There are in good agreement with calculated values in activity characteristics and there were small differences in dose distributions around the source in comparison between KAERI and old source.


Subject(s)
Academies and Institutes , Brachytherapy , Emergencies , Film Dosimetry , Korea , Nuclear Energy , Radius
6.
Korean Journal of Medical Physics ; : 30-36, 2009.
Article in Korean | WPRIM | ID: wpr-88370

ABSTRACT

The quality correction in the air kerma dosimetry for Ir-192 using farmer type ionization chambers calibrated by Co-60 quality is required. In this study we determined quality factor (kappa u) of two ionization chambers of PTW-N30001 and N23333 for Ir-192 source using dosimetric method. The quality factors for energy spectrum of microSelectron were determined as kappa u=1.016 and 1.017 for PTW-N30001 and N23333 ionization chambers respectively. We applied quality factors in air kerma dosimetry for microSelectron source and compared with reference values. As a results we found that the differences between reference air kerma rate and measured it with and without quality correction were about -0.5% and -2.0% respectively.


Subject(s)
Gamma Rays , Reference Values
7.
Korean Journal of Medical Physics ; : 43-50, 2009.
Article in Korean | WPRIM | ID: wpr-88368

ABSTRACT

The 4 bank mico-MLC (mMLC; Acculeaf, Direx, Isral) has been commissioned for clinical use of linac based stereotactic radiosurgery. The geometrical parameters to control the leaves were determined and comparisons between measured and calculated by the calculation model were performed in terms of absolute dose (cGy/100 MU). As a result of evaluating calculated dose for various field sizes and depths of 5 and 10 cm in water in the geometric condition of fixed SSD (source to surface distance) and fixed SCD (source to chamber distance), most of differences were within 1% for 6 MV and 15 MV x-rays. The penumbral widths at the isocenter were approximately evaluated to 0.29~0.43 cm depending on the field size for 6 MV and 0.36~0.51 cm for 15 MV x-rays. The average transmission and leakage for 6 MV and 15 MV x-rays were 6.6% and 7.4% respectively in single level of leaves fully closed. In case of dual level of leaves fully closed the measured transmission is approximately 0.5% for both 6 MV and 15 MV x-rays. Through the commissiong procedure we could verify the dose characteristics of mMLC and approximately evaluate the error ranges for treatment planning system.


Subject(s)
Radiosurgery , Silver Sulfadiazine , Water
8.
Korean Journal of Medical Physics ; : 56-62, 2008.
Article in Korean | WPRIM | ID: wpr-203475

ABSTRACT

In this study, we have investigated the dose characteristics of PTW-LinaCheck designed to detect output of medical LINAC and discussed clinical use of the detector. The reproducibility, linearity, and dose rate dependency of the dosimeter were measured for photons of 6 and 15 MV and the electrons of 4, 6, 9, 12, and 16 MeV. To know the error ranges of the measured data in daily output measurement, the response variations due to geometrical setup errors were measured. As a result of measurement, the error range from the geometrical setup and the reproducibility was less than +/-0.6% for given beam qualities in daily output measurement, where the errors from the linearity and the dose rate dependency were negligible. Finally, we concluded that the LinaCheck dosimeter has a good characteristics in terms of dose and setup convenience in daily output measurement. In addition we have shown an examples of clinical use of this dosimeter for measuring daily output more than 60 days.


Subject(s)
Dependency, Psychological , Electrons , Photons
9.
Korean Journal of Medical Physics ; : 113-119, 2008.
Article in Korean | WPRIM | ID: wpr-7199

ABSTRACT

In this study, we have estimated error of calculation results for 6-type RTP systems and investigated a toleration for error of the RTPs referenced from the evaluation items of AAPM Report-62. For this study, we have introduced the concept of 'normal dose rate(NDR)' and compared the results of experiment and calculation from RTPs at the same reference level. The results from all RTPs were satisfied at various field shapes and heterogeneous phantom materials except the surface irregularity.

10.
Korean Journal of Medical Physics ; : 131-138, 2008.
Article in Korean | WPRIM | ID: wpr-7196

ABSTRACT

We evaluated on the calculation accuracy of treatment planning system (TPS) with phantom having convex and concave surface. The TPS is Eclipse (Varian, USA) using both algorithms AAA and PBC for photon dose calculations. PBC algorithms have three corrections of Batho, modified Batho (M-Batho), and equivalent TAR (E-TAR). The field sizes were 10x10 cm2 and 20x20 cm2, and MLC-shaped fields for these fields. We measured doses at three depths 5, 10 and 15 cm in phantom of SSD=90 cm in the condition of inserted farmer chamber. For given conditions, we have calculated dose with these algorithms and compared them with measured doses. In AAA the calculated doses (dose/MU) were agreed to measured doses within +/-1% in flat and convex surface and were under estimated with -1.9% maximum in concave surface. In PBC the calculated doses were over estimated with +1.7% and +4.1% respectively in flat and convex surface and the differences were from -3.1% to +2.1% in concave surface. In comparison of criteria from AAPM and IAEA reports, and statistical analysis for these results, it is found that the AAA's results are in good agreement with measured values and the M-Batho's results are generally good agreed with measured values among PBC algorithms.


Subject(s)
Water
11.
Korean Journal of Medical Physics ; : 209-213, 2007.
Article in Korean | WPRIM | ID: wpr-213244

ABSTRACT

In this work we have measured the dose distribution and the percent depth dose of 20 MeV electron beam using the X-OMAT films in order to verify the effects of transverse magnetic field on high energy elecrtron beam in a phantom. The result shows about 30% increase of the percent depth dose at 4.5 cm depth under the transverse magnetic field of 1.5 Tesla at 7.5 cm depth. We have verified that these were in an agreement with other theoretical results.


Subject(s)
Magnetic Fields
12.
Korean Journal of Medical Physics ; : 221-225, 2007.
Article in English | WPRIM | ID: wpr-213242

ABSTRACT

Monte Carlo calculations were performed to demonstrate the dose modulation with dynamic magnetic fields in phantom. The goal of this study is to obtain the uniform dose distributions at a depth region as a target on the central axis of photon beam under moving transverse magnetic field. We have calculated the depth dose curves for two cases of moving magnetic field along a depth line, constant speed and optimal speed. We introduced step-by-step shift and time factor of the position of the electromagnet as an approximations of continuous moving. The optimal time factors as a function of magnetic field position were calculated by least square methods using depth dose data for static magnetic field. We have verified that the flat depth dose is produced by varying the speed of magnetic field as a function of position as a results of Monte Carlo calculations. For 3 T magnetic field, the dose enhancement was 10.1% in comparison to without magnetic field at the center of the target.


Subject(s)
Axis, Cervical Vertebra , Magnetic Fields , Magnets , Time Factors
13.
Korean Journal of Medical Physics ; : 111-115, 2005.
Article in Korean | WPRIM | ID: wpr-102314

ABSTRACT

We have designed the multi channel detector for the quality assurance of clinical photon beams. The detector was composed of solid phantom inserted by six plane-parallel ionization chambers at different depth. The chamber as a mini plane parallel chamber was made of carbon coated microfilms. In this study the electrical characteristics of the six chambers in the solid phantom were evaluated using 6 MV photon beam. The leakage currents were less than 0.5 pA, reproducibility was less than 0.5%, linearity was less than 0.5%, and dose rate effect was less than 0.7%. In addition the effect of dose variation from other chambers was estimated to maximum 0.8% approximately. The developed detector can be used for quality determination in output dosimetry or measurement of percentage depth dose approximately for clinical photon beam.


Subject(s)
Carbon , Electrodes
14.
Korean Journal of Medical Physics ; : 53-61, 2005.
Article in Korean | WPRIM | ID: wpr-187037

ABSTRACT

The quality factors for cylindrical ionization chambers for kV X-rays were determined by Monte Carlo calculation and measurement. In this study, the X-rays of 60~300 kV beam (ISO-4037) installed in KFDA and specified in energy spectra and beam qualities, and the chambers of PTW N23333 and N30001 were investigated. In calculations, the k(u) and k(Q,Q(0)) in IAEA dosimetry protocols were determined from the air kerma and the cavity dose obtained by theoretical and Monte Carlo calculations. It is shown that the N30001 chamber has a flat response of +/-1.7% in 110~300 kV region, while the response range of two chambers were shown to +/-3~4% in 80~250 kV region. From this work we have discussed dosimetry protocol for the kV X-rays and we have found that the estimation of energy dependency is more important to apply dosimetry protocol for kV X-rays.

15.
Korean Journal of Medical Physics ; : 97-103, 2005.
Article in Korean | WPRIM | ID: wpr-187031

ABSTRACT

The energy spectra and dose calculations were performed for secondary neutrons from a 24 MV LINAC using MCNPX code (V2, 4, 0). The energy spectra for neutrons and photons emitted from the LINAC head, and absorbed dose to water were calculated in water phantom. The absorbed doses calculated with Monte Carlo were 0.66~0.35 mGy/photon Gy at the surface to d=5 cm, and calculated with interaction data was 0.52 mGy/photon Gy at the depth of electron equilibrium in water. We have shown that this work can be applied to dose estimation of neutrons from high energy LINAC through the comparison of our results with other results.


Subject(s)
Head , Neutrons , Photons , Water
16.
The Journal of the Korean Society for Therapeutic Radiology and Oncology ; : 195-202, 1998.
Article in Korean | WPRIM | ID: wpr-185850

ABSTRACT

PURPOSE: In order to obtain basic data for treatment plan in radiosurgery, we measured small fields of 6 MV X-rays and compared the measured data with our Monte Carlo simulations for the small fields. MATERIALS AND METHODS: The small fields of 1.0, 2.0 and 3.0 cm in diameter were used in this study. Percentage depth dose (PDD) and beam profiles of those fields were measured and calculated. A small semiconductor detector, water phantoms, and a remote control system were used for the measurement. Monte Carlo simulations were performed using the EGS4 code with the input data prepared for the energy distribution of 6 MV X-rays, beam divergence, circular fields and the geometry of the water phantoms. RESULTS: In the case of PDD values, the calculated values were lower than the measured values for all fields and depths, with the differences being 0.3 to 5.7% at the depths of 2.0 to 20.0 cm and 0.0 to 8.9% at the surface regions. As a result of the analysis of beam profiles for all field sizes at a depth of 10cm in water phantom, the measured 90% dose widths were in good agreement with the calculated values, however, the calculated penumbra radii were 0.1 cm shorter than measured values. CONCLUSION: The measured PDDs and beam profiles agreement with the Monte Carlo calculations approximately. However, it is different when it comes to calculations in the area of phantom surface and penumbra because the Monte Carlo calculations were performed under the simplified geometries. Therefore, we have to study how to include the actual geometries and more precise data for the field area in Monte Carlo calculations. The Monte Carlo calculations will be used as a useful tool for the very complicated conditions in measurement and verification.


Subject(s)
Radiosurgery , Semiconductors , Water